Oakton® PC 510 Benchtop pH/Conductivity Meters
|

click to enlarge
|
|
EW-35610-10 Oakton® benchtop pH/conductivity/TDS meter with probes, 110/220 VAC |
| $919.00 / each (USD) | In stock. | | | |
|
|
 | Toggle easily between pH, conductivity, and TDS measurements |  | Dual display of pH, conductivity, or TDS plus temperature |  | Five-point calibration for pH and either conductivity or TDS |  | Conductivity and TDS autoranging chooses the best resolution for each measurement. |  | Previous calibration points, electrode slope and offset, and conductivity/TDS cell constants are stored in memory for quick and easy retrieva. |  | Measure pH and conductivity/TDS with separate probes to suit a wide variety of applications. |  | Features pH auto-buffer recognition, selectable °F/°C, HOLD, and water-resistant membrane keypad. | Meters include: Electrode holder arm, 110 VAC and 220 VAC power adapters, and instructions. Three-year warranty on meters only.
35610-10 also includes 59001-70 double-junction pH electrode with BNC connector & 35608-50 conductivity/TDS/temperature probe. |
|
| Specifications | | Range | TDS | 0.00 to 9.99, 10.0 to 99.9, 100 to 999 ppm; 1.00 to 9.99, 10.0 to 99.9, 100 to 200 ppt | | pH | -2.00 to 16.00 pH | | Temperature | 32 to 212°F (0 to 100°C) | | Conductivity | 0.0 to 19.99, 0 to 199.9, 0 to 1999 µS; 0 to 19.99, 0 to 199.9 mS | | Resolution | TDS | 0.01, 0.1, 1 ppm; 0.01, 0.1, 1 ppt | | Temperature | 0.1°F or °C | | Conductivity | 0.01, 0.1, 1 µS; 0.01, 0.1 mS | | Conductivity temp coefficient | Adjustable from 0.0 to 10% per °C | | Display | Dual LCD shows measurement plus temperature | | Power | 110 VAC | | Accuracy | Temperature | ±0.5°F or °C | | Calibration | Conductivity | up to 5 points (one point per range) | | TDS | up to 5 points (one point per range) | | Dimensions | 9"W x 2 3/8"H x 7"D | | pH | up to 5 points (pH 1.68, 4.01, 7.00, 10.01 and 12.45) | | Temperature | offset in 0.1° increments | | Operating temperature | 32 to 122°F (0 to 50°C) | | Resolution | pH | 0.01 pH | | Accuracy | pH | ±0.01 pH | | Conductivity | ±1% full scale | | TDS | ±1% full scale | | Temp compensation | Automatic or manual | | Conductivity cell constant | Fixed at k = 1.0 cm-1 | | Conductivity-to-TDS calibration values | Adjustable from 0.4 to 1.0 |
| Technical Appendix | |
|
|
|
Q: Does this pH meter measure the temperature of the solution and then measure the pH? Asked on 10/6/2009 by Anonymous Know the answer? Answer this question 1 answer A: This pH meter will measure and display the pH and temperature of the solution at the same time. If you have additional questions, please contact our Application Specialists at the number listed on our website. Answered on 10/8/2009 by Jaime from Technical Support
Q: will this ph probe work effectively in a solution of soil:water at the ratio 1:2.5, which is being stirred? Asked on 9/15/2009 by Anonymous from suffolk Know the answer? Answer this question 1 answer A: The probe with his meter is fine for measuring the pH in a water and soil mixture. You should clean the probe after taking measurments with cleaning solution 00653-00. Answered on 9/15/2009 by Aaron from Tech Support
Q: Does the pH electrode work well with Tris buffers or can suffer any kind of damage with these buffers? Asked on 7/9/2009 by Anonymous Know the answer? Answer this question 1 answer A: The pH probe that comes with this meter is catalog number 59001-70, which will work well in TRIS buffer solutions. Answered on 7/9/2009 by Jen from Technical Support
Q: Are there power supply cords available for model # 59003-00 Asked on 6/10/2009 by Anonymous Know the answer? Answer this question 1 answer A: The replacement 110 VAC power supply is sku 35621-50. Answered on 6/10/2009 by Staff Expert from Cole-Parmer
|
[2 of 2 customers found this review helpful] pH probe works but has some issues By Dawgfan96 from Seattle, WA on 5/7/2009 Pros: Accurate, Durable, Easy To Use Best Uses: In The Lab, Research Describe Yourself: Academic/Professional Researcher Cole-Parmer Response: We apologize for your difficulties and appreciate you bringing this to our attention--thank you! Recently, we upgraded the connectors on the electrodes as they are also used with portable instruments which require additional durability and watertight connections. The rubber sleeves on the connectors will slide as needed to allow attachment to our benchtop meters. Unfortunately when the upgrade was made, we did not communicate this change in the corresponding manual & literature. Our engineering team is aware of the issue and will take appropriate action. Please contact our technical support team for calibration assistance--they can walk you thru a calibration.
|
|
| |
 |
|
|
|