Wavelength Noise±0.002 A at 0.04 A, ±0.02 A at 2.0 A and 546 nm
Wavelength Accuracy±2 nm
Wavelength Reproducibility±2 nm
Pathlength (mm)10 to 100
Min Photometric - Absorbance (A)-0.3
Max Photometric - Absorbance (A)2.5
Photometric Noise±0.002 A at 0.04 A
Photometric Drift±0.005 A/h at 0.04 A and 546 nm after 60 min.
Photometric Accuracy±0.01 A at 1.0 A and 546 nm
DetectorDiode Array
Concentration±2500
OutputUSB
Display TypeTouch screen LCD
Width (cm)21.2
Height (cm)12
Depth (cm)42.2
Width (in)8
Height (in)4 3/4
Depth (in)16
Power (VDC)12
Power (VAC)100 to 240
Power (Hz)50/60 Hz
DescriptionGenova Bio UV/Visible Diode Array Scanning Spectrophotometer
Includes
Micro-cuvette holder, USB memory stick, and power adapter with three plugs (US, EU, and UK).
Key Features
Simultaneous measurement of all wavelengths increases productivity
Fast scan speed—as short as 3 seconds
Comapct and lightweight
Multiple measurement modes for broad application compatibility
Sample chamber lid can be left open—ideal for tall test tubes
Color touch screen and user-friendly interface make setup easy
Dedicated to life science applications—preprogrammed for DNA, RNA and protein analysis
Narrow spectral bandwidth of 3 nm identifies peaks for determining nucleic acid purity
Wavelength range: 198 to 800 nm
Spectral bandwidth is 3 nm
More About this Item
The multi-element detectors on these scanning spectrophotometers measures the entire wavelength range at once, reducing wait times. Measurement modes include photometrics, concentration, optical density (O.D. at 600 nm), quantitation, spectrum scanning, and kinetics. Each dataset contains readings from the complete spectrum, making the data readily available for future analysis.
The orientation of the detector shields the array from stray light—perform measurements with the lid open, exchange samples quickly, and decreases user errors in results. The large color touch screen is easy to read, and the graphical interface is intuitive. The USB port on the rear connects with a printer (order separately), while the USB on the front accepts flash drives for data storage, quantitation curve uploads, and firmware updates.
Use the preprogrammed nucleic acid and protein measurement modes to calculate concentration levels. The nucleic acids measurement mode quantifies the concentration and purity of dsDNA, sDNA, RNA, and oligonucleotides. A simple touch of a button can check the purity of the nucleic acids. Calculate the protein concentration using the protein measurement mode. Preprogrammed methods for measuring Bradford, Lowry, Biuret, and Bicinchoninic Acid (BCA) assays minimize errors.
Our specialists are here to help you find the best product or part available for your application. Call or Email us and we will make sure you get the right product or part for the job.