Masterflex L/S Digital Drive, 600 rpm; 115/230 VAC

Representative image only.
Masterflex Item # EW-07522-20
Precision digital drive offers a maintenance-free high-accuracy motor and a graphical interface – makes setup and operation easier than ever!
  • Flow rate 0.001 to 3400 mL/min with L/S tubing. Flow rate depends on drive rpm and tubing size.
  • Graphical LCD shows four operating modes continuous run, timed dispense, copy dispense, volume dispense. View pump performance continually.
  • Dispense by Volume in mL (0.001 to 9,999 mL), L, or gal; Copy—1 to 9,999 dispense cycles; or Time—time interval of on/off cycles (1 sec to 99 hr, 59 min, 59 sec per cycle)
  • Antidrip function ensures dispensing accuracy
  • Menu-driven interface lets you select one of seven languages for programming and operation

PRODUCT OPTIONS

$2,212.00USD / Each

$133.00USD / Each

for an additional
$133.00USD / Each
Time to recalibrate? Click on the calibration service to view details, then order.
Available with MasterflexLive Remote Pump Control and Monitoring

SPECS & DESCRIPTION

Specifications & Description

  • Pump SeriesL/S
  • Control TypeDigital Variable Speed
  • Min RPM (rpm)0.1
  • Max RPM (rpm)600
  • SpeedVariable
  • Speed Control±0.1% (0.1 rpm at 600 rpm)
  • Max Flow Rate (mL/min)3400
  • Min Flow Rate (mL/min)0.006
  • Reversible MotorYes
  • Digital DispensingYes
  • No. of Pump Heads Accepted2
  • Type of Pump Heads AcceptedStandard, Easy-Load, Easy-Load II, Easy-Load 3, High Performance, Cartridge, Multichannel, PTFE Tubing, PTFE Diaphragm
  • Motor (watts)75
  • Drive IP ratingIP 33
  • Power (VAC)90 to 260
  • Power (amps)2.2
  • Power (Hz)50/60
  • Drive Length (in)10
  • Drive Width (in)8.5
  • Drive Height (in)8.5
  • Drive Length (mm)254
  • Drive Width (mm)216
  • Drive Height (mm)216
  • Open-Head SensorNo
  • Housing MaterialABS Plastic and Aluminum
  • Motor hp1/10
  • Remote I/O0 to 20 mA, 4 to 20 mA, 0 to 10 V, start/stop, direction, and prime via contact closure
  • DescriptionL/S Digital Drive, 600 RPM, 115/230 VAC 0
Two year warranty

Includes

6-ft (1.8-m) line cord with IEC 320/CEE22 socket. These models are shipped with country specific plug/cord set; please specify ultimate destination when ordering.

Key Features

  • Flow rate 0.001 to 3400 mL/min with L/S tubing. Flow rate depends on drive rpm and tubing size.
  • Graphical LCD shows four operating modes continuous run, timed dispense, copy dispense, volume dispense. View pump performance continually.
  • Dispense by Volume in mL (0.001 to 9,999 mL), L, or gal; Copy—1 to 9,999 dispense cycles; or Time—time interval of on/off cycles (1 sec to 99 hr, 59 min, 59 sec per cycle)
  • Antidrip function ensures dispensing accuracy
  • Menu-driven interface lets you select one of seven languages for programming and operation
  • Brushless, maintenance-free motor offers ±0.1% speed control accuracy with a 6000 to 1 turndown ratio
  • Digital drive features a maximum speed of 600 rpm

More About this Item

These precision drives offer analog remote control options and simple programming—ideal for automated process applications. Drives feature a programmable dispense interval that lets you set the delay between dispense cycles for convenient, automated dispensing. Programmed calibration ensures dispense and display accuracy—drive stores one calibration value per tubing size.
Program the drive via the simple membrane keypad; keypad lock/unlock feature protects settings. Motor is reversible so you can purge tubing before or after pumping and pump fluid in either direction.

PART OF A COMPLETE SYSTEM – GET EVERYTHING YOU NEED
Masterflex 77921-70 L/S Digital Pump System with Easy-Load II Pump Head, 600 RPM, 115/230 VAC 77921-70
The convenient modular design of Masterflex Pump Systems (Pump Head + Tubing + Drive) makes Masterflex extremely versatile and cost-effective.

ACCESSORIES

REVIEWS

REVIEW SNAPSHOT®

by PowerReviews
Masterflex L/S Digital Drive, 600 RPM, 115/230 VAC 0
 
3.0

(based on 3 reviews)

67%

of respondents would recommend this to a friend.

Pros

No Pros

Cons

No Cons

Best Uses

No Best Uses
    • Was this a gift?:
    • No (3)

Reviewed by 3 customers

Displaying reviews 1-3

Back to top

 
4.0

Good dependable and accurate.

By 

from Dallas Texas

Verified Buyer

Pros

  • Very Dependable

Cons

  • Expensive

Best Uses

  • Ls Easy-load Ii Head

Comments about Masterflex L/S Digital Drive, 600 RPM, 115/230 VAC 0:

Metering chemicals into production vats

  • Was this a gift?:
  • No

(1 of 2 customers found this review helpful)

 
1.0

Pumps are great until they fail...and they will fail

By 

from Berkeley, CA

Comments about Masterflex L/S Digital Drive, 600 RPM, 115/230 VAC 0:

We have purchased a number of these pumps over a couple year period and now have more than half a dozen of them. Every single one of them has now failed with the exact same error code and had to have the board replaced. Which is especially problematic because a few of them failed mid experiment and after the warranty expired. I think if a company knows their product is flawed they should recall it instead of waiting until the warranty expires and forcing the customer to pay for a repair on a defective device.

  • Was this a gift?:
  • No
 
4.0

Makes life easier!

By 

from Carlsbad, CA

Verified Reviewer

Comments about Masterflex L/S Digital Drive, 600 RPM, 115/230 VAC 0:

We are overall satisfied with this product. Although a bit expensive for what it is, in the long run it is worth it and is good value. It is so much easier to use this device then to manually pipet large amounts of tubes. The interface is simple and easy to use. The only down side is that the calibration process takes a bit too long, and we often have to estimate then measure ourselves. But for aliquoting the same volume it seems to be very accurate.

  • Was this a gift?:
  • No

Displaying reviews 1-3

Back to top

Q&A

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

8 Questions | 8 Answers
Displaying questions 1-8
  • Q:

    How many easy load ii attachments can one pump handle?
    Asked on 9/27/2016 by fish from Wisconsin

    1 answer

    • STAFF EXPERT

      A:

      The maximum number of pump heads that should be mounted to this pump drive is a total of 2 pump heads.

      Answered on 10/2/2016 by Jose from Technical Support
  • Q:

    Can you clarify the power requirement for this pump? We are having it run on a solar panel/battery system with an inverter and I want to make sure the loading is only 75W. The 2.2A at 115VAC rating is what has me worried, but that is misleading because current and voltage are not in phase correct(i.e. P does not equal 2.2A times 115V)?
    Asked on 7/11/2016 by Ryan from San Diego, CA

    1 answer

    • STAFF EXPERT

      A:

      Supply voltage limits: all models 90 to 260 Vrms @ 50/60 Hz
      (Universal Input) Single Phase Only
      Current, max.: all models 1.8A @ 115 Vrms, or 1.1A @ 230 Vrms

      Answered on 8/12/2016 by Dan C from Technical Support
  • Q:

    Will this pump work with the hand-held remote controller, part number 07528-80, which is available for the Variable-Speed Console Drive, HV-07258-10?
    Asked on 4/19/2016 by CuriousCat from Bellevue, WA

    1 answer

    • STAFF EXPERT

      A:

      This pump will not work with that specific hand-held remote because the PIN connections would be different. You could however utilize the hand-held remote part number 07523-97 with this drive.

      Answered on 4/20/2016 by Rishi from Technical Support
  • Q:

    we are using this pump remote, controlling it through the DB9 connector. Is there a way to count the motor rotations?
    Is something known about mounting a hall sensor to the head? Can you please advise what the best way is to go?
    Thank you in advance,
    Best regards,
    Asked on 1/6/2016 by msmeulers from delft, the netherlands

    1 answer

    • STAFF EXPERT

      A:

      Unfortunately, this unit does not allow for a way to count motor rotations remotely. The RPM is only available to be displayed in continuous mode on the screen. This unit has a DB25 connector mounted on the back for remote outputs. Pin 1 represents voltage input and pin 14 represents voltage output.

      Answered on 1/6/2016 by Katie from Technical Support
  • Q:

    Hi, how often should I do the calibration?
    Asked on 10/25/2015 by Ying from Princeton, NJ

    1 answer

    • STAFF EXPERT

      A:

      It would depend on how often you use the pump. You should calibrate anytime the application changes to receive optimum accuracy. You should calibrate when the tubing is changed as well. If you see an issue with the accuracy, the pump should be calibrated.

      Answered on 10/26/2015 by Danielle from Technical Support
  • Q:

    Can the break - in be done with water or do we have to use actual product?
    Asked on 7/13/2015 by Beverly from Fl

    1 answer

    • STAFF EXPERT

      A:

      You can use water or just run the tubing dry.

      Answered on 7/13/2015 by Anthony from Technical Support
  • Q:

    pump was flammable rated (NFPA Class I, Div 2
    Asked on 7/8/2015 by gan from greenville, Sc

    1 answer

    • STAFF EXPERT

      A:

      This pump is not listed for hazardous locations that need to meet Class 1 Div 2 standards.

      Answered on 7/8/2015 by Danielle from Technical Support
  • Q:

    Hi

    I used to own the older model of the L/S pump WU-07523-50. My application requires the pump connecting to a timed power outlet. With the old model, when the power is cut and resumed the drive automatically turns back on with the preset speed.

    However, recently when I replace it with this newer model WU-07522-30, the drive will not resume running after the power is cut off and resumed.

    Is there a setting or workaround where I can control the pump with a timed outlet like the older model?

    Thanks!!
    Asked on 1/15/2015 by Brian from United States

    1 answer

    • STAFF EXPERT

      A:

      Power cycling this drive with a timer is not recommended. I would suggest using the internal batch dispensing options to set the delay between cycles.

      Answered on 1/16/2015 by Jay from Technical Support