Cole-Parmer Four-Element Flame Photometer 02655-15

Representative image only.
Cole-Parmer Item # EW-02655-15
Provide accurate sodium, potassium, calcium, and lithium determinations
  • Features additional calcium channel and lithium internal standard

$14,600.00USD / Each


Specifications & Description

  • Rangeppm: all 1.00 to 9.99 ppm, 10.0 to 199.9; mmol/L: as above except Li only 1.00 to 5.00
  • Specificity<0.5% interference when concentrations are equal to test sample concentrations
  • SensitivityNa & K = 0.5 ppm, Li = 2 ppm, Ca = 5 ppm
  • LinearityLess than 1% at midpoint with 3 ppm K set at 100
  • ReproducibilityLess than 0.5% CV
  • Reading StabilizationVariable software command
  • Aspiration Rate3 to 6 mL/min
  • Air Supply6 L/min at 15 psi; oil and moisture free
  • Display3.5-digit LED; 12.5 mm (1/2") H
  • Fuel SupplyHigh-grade propane/butane, or propane/butane mixture regulated at approximately 30 psi; natural gas - only with single-channel photometer
  • Recorder OutputTwo outputs: 50 mV to 7.5 V via software command
  • Height (in)20
  • Depth (in)13 1/2
  • Width (in)15 3/8
  • Height (cm)51
  • Depth (cm)34.5
  • Width (cm)39
  • Power (VAC)120 / 240
  • Power (Hz)50/60
  • DescriptionFour-Element Flame Photometer 02655-15


one each of Na+, K+, Ca+, Li+ filters, one each of 1000 ppm Na+, Ca+, K+, Li+ standards in a 100-mL bottle, two meters each of gas and air tubing with connectors, RS-2

Key Features

  • Features additional calcium channel and lithium internal standard

More About this Item

This enhanced four-element flame photometer offers the same great features found on our dual-channel flame photometer, plus a calcium channel and direct measurement of Ca+ with potassium, which are two vital elements for soil analysis. Unit allows you to measure calcium in combination with sodium, and also with lithium without an internal standard.
Use this flame photometer as a detector in combination with multi-channel analyzers from Bran & Luebbe, Skalar, and Lachat. This system includes its own protocols for calibration, and features a special mode allowing you to set up the flame photometer within these protocols. User-selectable signal averaging is designed into the system allowing you to use these protocols along with analog output. Features include automatic ignition sequence, lithium internal standard, monitoring and control software, dual-channel display, and independent calibration of each channel. An automatic gas shutoff mechanism activates if the flame is accidentally extinguished, or if air pressure drops.
Order optional Automatic Dilutor 02657-00 which accurately dilutes your samples and standards; order optional Autosampler 02657-25 for a fully automatic analysis solution for your longer sampling runs.




Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

2 Questions | 2 Answers
Displaying questions 1-2
  • Q:

    Is this instrument suitable for Clinical applications?
    Asked on 4/6/2016 by paulo

    1 answer



      This item is not recommended for any type of clinical or medical use.

      Answered on 4/6/2016 by Katie from Technical Support
  • Q:

    We want to purchase CFR 21 compliance software with Flame photometer, so pl. suggest model with detail.
    Asked on 1/6/2016 by Sunil Choksi from Indore, India

    1 answer



      Unfortunately, this software package is not CFR 21 compliant. This is the only software available for the flame photometers we offer.

      Answered on 1/7/2016 by Katie from Technical Support
Displaying questions 1-2

Do you have a question about this product? 


Need Help with this product?

Our specialists are here to help you find the best product or part available for your application. Call or Email us and we will make sure you get the right product or part for the job.

Call Us At1-800-323-4340
Live chat