This item is now discontinued. Please browse similar items

Hach 51800-10 sensION 5 Waterproof Conductivity Meter

Representative image only.
Hach Mfr # 51800-10 – Item # EW-19604-00
Chemical-resistant, dustproof, and waterproof meter for your most demanding applications
  • Ultra-fast response—get your result within seconds
  • IP67-rated housing—waterproof, rugged, and it floats!
  • Large backlit LCD with anti-glare coating

SPECS & DESCRIPTION

Specifications & Description

  • Display typeLCD
  • Conductivity Resolution0.1 or 1 μS; 0.01 or 0.1 mS
  • Conductivity Accuracy±0.5% FS
  • TDS Accuracy±0.5% FS
  • Min Salinity0 to 43 ppt
  • Salinity Resolution0.1 ppt
  • Salinity Accuracy±0.1 ppt
  • Temperature Accuracy±0.5 °C
  • Temperature compensationAutomatic
  • InterfaceRS-232
  • WaterproofYes
  • Width (cm)8.73125
  • Height (cm)21.192998
  • Width (in)3 7/16
  • Length (in)1 21/32
  • Height (in)8 11/32
  • DescriptionsensION 5 Waterproof Conductivity Meter

Includes

sensION four-band conductivity cell (19604-50), beakers, and batteries.

Key Features

  • Ultra-fast response—get your result within seconds
  • IP67-rated housing—waterproof, rugged, and it floats!
  • Large backlit LCD with anti-glare coating

More About this Item

Four-band conductivity cell (included) covers the entire measuring range of the meter—reduces the effects of polarization. Stores up to 99 data points; download data to a printer or computer via the docking station 59005-50 (order separately).

ACCESSORIES

REVIEWS

Q&A

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

2 Questions | 2 Answers
Displaying questions 1-2
  • Q:

    How often should the meter be calibrated?
    Asked on 5/31/2011 by Anonymous

    1 answer

    • STAFF EXPERT

      A:

      The frequency of calibration will depend on your accuracy needs and frequency of use. If accuracy is of major concern, I suggest calibrating once a day, before taking your measurements. If your accuracy requirements are not as strict, you could easily move to a once a week calibration schedule.

      Answered on 6/3/2011 by Josh from Technical Support
  • Q:

    How do i calibrate the salinity meter? Is there buffer solution i can purchase?
    Asked on 7/8/2010 by Anonymous

    1 answer

    • Staff Reviewer

      A:

      Yes, there are conductivity and salinity calibration standards. The most common conductivity/salinity standards are part numbers 00653-47 (215 ppm NaCl), 00653-18 (745 ppm NaCl). There are others that may work better for the range you are working in though. Please contact our technical application specialists for additional assistance (contact information can be found in the "contact us" section on the web page).

      Answered on 7/9/2010 by Jen from Technical Support
Displaying questions 1-2