Cole-Parmer Visible Spectrophotometer; 335 to 1000 nm Wavelength Range; Analog Output

Representative image only.
Cole-Parmer Item # EW-83055-10
  • Economical, non-scanning, general-purpose laboratory spectrophotometers

PRODUCT OPTIONS

$1,378.50USD / Each

BUY MORE & SAVE

  • 1-4$1,378.50USD / Each
  • 5-9$1,240.65USD / Each
  • 10-19$1,171.73USD / Each
  • 20+$1,102.80USD / Each

$265.80USD / Each

for an additional
$265.80USD / Each
Time to recalibrate? Click on the calibration service to view details, then order.

SPECS & DESCRIPTION

Specifications & Description

  • Stray Light<0.5%T at 340 nm, 400 nm
  • Source LampTungsten-Halogen
  • Bandwidth20 nm
  • Min Wavelength (nm)335
  • Max Wavelength (nm)1000
  • Wavelength Accuracy±2.0 nm
  • Min Photometric - Transmittance (%T)0
  • Max Photometric - Transmittance (%T)125
  • Min Photometric - Absorbance (A)0
  • Max Photometric - Absorbance (A)2
  • DetectorSolid-state
  • OutputUSB
  • Display TypeLCD (4-digit)
  • DescriptionVisible spectrophotometer; 335 to 1000 nm wavelength range; analog output
Two year warranty

Key Features

  • Economical, non-scanning, general-purpose laboratory spectrophotometers

More About this Item

These spectrophotometers suit a variety of general-purpose and quality control applications—ideal for biochemistry, environmental chemistry, food testing, and water testing labs. They combine value and simplicity with the accuracy of much more expensive instruments.

Auto-zeroing saves time between samples and reduces potential errors. The wavelength adjustment knob features end-of-range stops to maintain calibration and prevent damage. Built-in filters automatically switch to accommodate wavelength setting—eliminating manual adjustment and ensuring accuracy.

What’s included: twelve 10-mm diameter test tubes cuvettes, cuvette adapter, dust cover, and 7-ft (2.1-m) cord with a US standard plug. For 230 VAC operation, order the optional 83055-25 power cable.

ACCESSORIES

REVIEWS

Q&A

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

11 Questions | 11 Answers
Displaying questions 1-10Previous | Next »
  • Q:

    What is the resolution of the absorbance measurement?
    Asked on 1/30/2017 by Andreas from Fyn

    1 answer

    • STAFF EXPERT

      A:

      The resolution is 0.000-1.999

      Answered on 2/21/2017 by Anthony from Technical Support
  • Q:

    Can this item be shipped to Denmark?
    Asked on 1/30/2017 by Andreas from DK

    1 answer

    • STAFF EXPERT

      A:

      This should be fine to export to Denmark, but our International Department would have to confirm.

      Answered on 1/31/2017 by Eric from Technical Support
  • Q:

    What is the footprint of this unit?
    Asked on 12/8/2016 by Vicki

    1 answer

    • STAFF EXPERT

      A:

      The dimensions are 408mm Width x 308mm Depth x 185mm Height.

      Answered on 12/8/2016 by Nicole from Technical Support
  • Q:

    What type of cord/connector is needed to connect the spec to a computer in order to use the software?
    Asked on 7/21/2016 by Cindy from Chestertown, MD

    1 answer

    • STAFF EXPERT

      A:

      If you purchase the PC application software item number EW-83055-21; the RS232 cable to connect to a computer is included.

      Answered on 8/3/2016 by Dan C from Technical Support
  • Q:

    Do you sell replacement lamps for the EW-83055-10 spec?
    Asked on 5/11/2016 by bdpia

    1 answer

    • STAFF EXPERT

      A:

      Yes, we offer the replacement halogen bulb as item 00141FY.

      Answered on 5/11/2016 by Andrew from Technical Support
  • Q:

    I am trying to test the content of iron in a natural occurring magnesium chloride brine. we need a concentration of 1ppm or less. which instrument would you recommend, and would the brine need to be diluted even if it is really clear?
    Asked on 10/1/2015 by Lisa from austin, tx

    1 answer

    • STAFF EXPERT

      A:

      Generally testing for iron content with use of a spectrometer require reagents/test kits. We do not have procedures on file to reference. Please contact our technical application specialists for additional assistance (contact information can be found in the "contact us" section on the web page).

      Answered on 10/1/2015 by Jen from Technical Support
  • Q:

    Does this unit accept 16 mm round test tube style cuvettes?
    Asked on 10/30/2013 by Shane

    1 answer

    • STAFF EXPERT

      A:

      No, it will not accept 16 mm diameter cuvettes. This unit comes with a sample holder to accept 10 mm diameter cuvettes, or standard square cuvettes with 10 mm pathlength.
      Please see the 83054-00 spectrometer instead, and note the 83054-51 adapter will be needed to accept a 16 mm diameter test tube cuvette.

      Answered on 10/31/2013 by Jen from Technical Support
  • Q:

    what are the main objectives of using PC software for this model? Will it work without PC software?
    Asked on 7/24/2013 by akakios from Georgia

    1 answer

    • STAFF EXPERT

      A:

      The 83055-10 will operate without the software. The software allows for easy data collection and application expansion to Standard Curve and Kinetics, Absorbance and % Transmittance/Concentration data can be easily exported to Microsoft Excel for further processing and analysis.

      Answered on 7/25/2013 by Polo from Technical Support
  • Q:

    Can this unit test solid materials?
    Asked on 7/11/2013 by TD

    1 answer

    • STAFF EXPERT

      A:

      This unit is meant for use with liquid samples only.

      Answered on 7/12/2013 by Polo from Technical Support
  • Q:

    I'm trying to test water containing Ni, Cr, Zn, Cd, Fluorides, Sulphates etc for a metal finishing industry. Can I do so using this instrument?
    Asked on 10/27/2010 by Sunny from Tampa, Fl

    1 answer

    • Staff Reviewer

      A:

      Generally testing for metals and with use of a spectrometer require reagents/test kits. We do not have procedures on file to reference. Please contact our technical application specialists for additional assistance (contact information can be found in the "contact us" section on the web page).

      Answered on 10/29/2010 by Jen from Technical Support
Displaying questions 1-10Previous | Next »

Do you have a question about this product? 

CUSTOMER SUPPORT

Need Help with this product?

Our specialists are here to help you find the best product or part available for your application. Call or Email us and we will make sure you get the right product or part for the job.

Call Us At1-800-323-4340
Live chat