Seward Stomacher 400C Lab Blenders, programmable, 110VAC

Representative image only.
Seward Mfr # 0400/001/AJ – Item # EW-36850-04
Blend and homogenize within sealed processing bags – eliminating cross-contamination and cleaning between samples
  • Improved organism recovery with the patented curved shape of the irculator paddles over older square paddle designs
  • No risk of cross contamination between samples
  • Durable and reliable
  • Control panel with programmable settings or ‘hands-free’ auto-run
  • Countdown of processing time

PRODUCT OPTIONS

$6,140.00USD / Each

SPECS & DESCRIPTION

Specifications & Description

  • Bag Capacity (mL)400
  • Power (VAC)110
  • Door TypeSolid
  • Timer (minutes)99
  • Width (in)13 1/4
  • Height (in)11
  • Depth (in)13 1/4
  • Width (cm)33.7
  • Height (cm)27.9
  • Depth (cm)33.7
  • DescriptionLab Blenders, programmable, 110VAC

Key Features

  • Improved organism recovery with the patented curved shape of the irculator paddles over older square paddle designs
  • No risk of cross contamination between samples
  • Durable and reliable
  • Control panel with programmable settings or ‘hands-free’ auto-run
  • Countdown of processing time
  • Removable door for easy cleaning of paddle chamber
  • 5 year warranty and 10 year service support

More About this Item

Stomacher 400C Meeting the food safety and quality demands of today’s microbiologists.

Whatever method you use to enumerate and identify bacteria, yeasts and moulds in your laboratory, thorough reproducible homogenisation of the sample is the first and most critical step.

Numerous studies have demonstrated the Circulator enhancement as a positive benefit to pathogen testing procedures in traditional microbiology and PCR based techniques.

As the sample is crushed during operation, the patented curved shape of the Circulator paddles create a washing effect driving more organisms in suspension thereby providing a better homogenised sample for analysis.
The Stomacher® 400 Bag Range

Caters for both the Circulator bag (round-bottom) designed bags, and the Classic (square-bottomed) for the older square paddle designed Stomachers.

Warning: This product is not approved or intended for, and should not be used for medical, clinical, surgical or other patient oriented applications.

REVIEWS

Q&A

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

1 Question | 1 Answer
Displaying question 1
  • Q:

    What method would you suggest for calibrating a stomacher, specifically the Stomacher 80 Biomaster? Thank you.
    Asked on 8/21/2009 by Kate from Bridgend

    1 answer

    • STAFF EXPERT

      A:

      Since they are blenders and not devices used to measure or analyze, stomachers don't really require much of a calibration. The only things that can be calibrated or checked for calibration are the timer and the speed. You can calibrate or check to be sure the time is accurate against a stopwatch and speed can be checked using an optical tachometer such as the EW-87702-00. Both of which are considered in spec if they are within 5%.

      Answered on 8/26/2009 by Josh from Technical Support
Displaying question 1

Do you have a question about this product? 

CUSTOMER SUPPORT

Need Help with this product?

Our specialists are here to help you find the best product or part available for your application. Call or Email us and we will make sure you get the right product or part for the job.

Call Us At1-800-323-4340
Live chat