Thermo Scientific Orion 8157BNUMD Ultra pH/ATC Probe, DJ/Epoxy; BNC/8-Pin DIN

Representative image only.
Thermo Scientific Orion Mfr # 8157BNUMD – Item # EW-05718-34
2-year replacement warranty
  • Builds on the ROSS® technology for greater stability, relability, and renowned accuracy and temperature response
  • Less long term drift saves time from recalibrating
  • Double junction
  • Epoxy body
  • 8-Pin DIN temperature compensation

PRODUCT OPTIONS

$538.20USD / Each
$598.00USD / Each

SPECS & DESCRIPTION

Specifications & Description

  • Electrode TypeDouble junction
  • Electrode Body MaterialEpoxy
  • Reference JunctionGlass fiber
  • Reference ElectrolyteROSS
  • Min pH (pH)0
  • Max pH (pH)14
  • Accuracy±0.01 pH
  • Temperature CompensationYes
  • ATC Connector8-pin mini-DIN
  • Connector8-pin mini-DIN
  • Diameter (mm)12
  • Length (mm)120
  • Cable Length (m)1
  • DescriptionUltra pH/ATC Probe, DJ/Epoxy; BNC/8-Pin DIN

Key Features

  • Builds on the ROSS® technology for greater stability, relability, and renowned accuracy and temperature response
  • Less long term drift saves time from recalibrating
  • Double junction
  • Epoxy body
  • 8-Pin DIN temperature compensation
  • BNC connection

More About this Item

The ROSS Ultra stability gives you ±0.01 pH accuracy even over a long period of time. Using a nonmetallic reference system, the liquid-liquid equilibration occurs almost instantaneously, saving time plus ensuring accurate results. Complex, delicate samples, such as biological media, foods, and pharmaceuticals, measure easily without metallic contamination of the sample or excessive cleaning. ROSS Ultra electrodes are refillable—easily accommodating a change in filling solution for samples sensitive to potassium or chloride ions. Choose ROSS Ultra pH/ATC Triodes for applications requiring built-in temperature compensation.

Use with Orion Star Series pH meters.

ACCESSORIES

REVIEWS

Q&A

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

5 Questions | 5 Answers
Displaying questions 1-5
  • Q:

    will this unit works with orion 230A+
    Asked on 2/14/2017 by germs

    1 answer

    • STAFF EXPERT

      A:

      This probe has a BNC and a 8-Pin DIN connector and will be compatible with meters that have those inputs.

      Answered on 2/14/2017 by Mike from Tech
  • Q:

    what is the accuracy/reproducibility of the pH 8157BNUMD probe?
    Asked on 6/22/2016 by Tim

    1 answer

    • STAFF EXPERT

      A:

      Here are some of the characteristics of the PH electrode:
      Unmatched response in samples with
      varying temperatures. Stable to 0.01
      pH within 30 seconds in 6.86 buffer
      between 25 °C and 75 °C
      • Accuracy within 0.03 pH when using
      automatic temperature compensation
      • Precision of 0.01 pH
      • Reliable, reproducible results
      • Double junction design prevents sample
      contamination. Filling solution can be
      changed when potassium or chloride in
      the sample is not desired

      Answered on 8/1/2016 by Dan C from Technical Support
  • Q:

    Hi,
    I need to now the acceptable slope range that indicates the probe is performing 'good". For example, other probes i use have the following;
    95-105% (good)
    <95% (probe needs a clean)
    >105% change the probe).

    Thanks,
    Martina
    Asked on 11/13/2014 by Tina from Auckland, NZ

    1 answer

    • STAFF EXPERT

      A:

      The slope should be between 92% and 102%.

      Answered on 11/14/2014 by Jay from Technical Support
  • Q:

    Is this electrode good for measuring nanoparticle (ceramic and metallic) solutions? If not, is there a good electrode that works with Orion SA720 model.
    Asked on 5/2/2011 by lsudhee from Davis, CA

    1 answer

    • STAFF EXPERT

      A:

      Yes, this will work for your solution.

      Answered on 5/3/2011 by Mike from Tech Support
  • Q:

    I need to know the minimum measurable volume with this probe.
    Asked on 6/2/2010 by Aaron from MA

    1 answer

    • STAFF EXPERT

      A:

      The minimum readable volume will depend on the container the sample is in. In order to get a reading there needs to be enough fluid to cover the bulb at the end of the probe (~5mm).

      Answered on 6/4/2010 by David from Tech Support