  • Q:

    what is the interference with chloride ? can it be used in an effluent with 2000 mg/l Cl?
    Asked on 8/14/2015 by caro from United States

    1 answer

    • STAFF EXPERT

      A:

      This is a nitrate ion-selective electrode. The interferences with this nitrate ISE are C1O4 -1, I-1, Cn-1, BF4 -1, Cl-1, and NO2 -1.

      Answered on 8/14/2015 by Danielle from Technical Support
  • Q:

    Does the "nitrate" probe that goes with the YSI Pro Plus multi-parameter, measure nitrate or nitrate-N? The calibration solution is nitrate-N, so I wasnt sure.
    Asked on 9/27/2012 by NA

    1 answer

    • STAFF EXPERT

      A:

      This probe measures Nitrate (NO3-).

      Answered on 10/1/2012 by Alyssa from Tech Support
  • Q:

    Are you sure that temperature resolution is 0.01C?
    Asked on 10/5/2010 by K from E

    1 answer

    • Staff Reviewer

      A:

      This part number represents a probe, which does not have a display.
      Please contact our technical application specialists for additional assistance and specify which meter you are interested in (contact information can be found in the "contact us" section on the web page).

      Answered on 10/6/2010 by Jen from Technical Support
  • Q:

    Is this a complete unit for measuring nitrate ions? Or, do I need to purchases separately a meter to which this probe is attached? If the latter is true, what meter (catalog #) do you recommend for a low-cost system?
    Asked on 7/2/2010 by Joanne Belovich from Cleveland Ohio

    1 answer

    • STAFF EXPERT

      A:

      This part number is just the electrode. The meter and standard/solutions should be purchased separately. This Nitrate electrode is for the 59352-00 YSI multiparameter meter. For a low cost system you can purchase the 27504-22 Nitrate ISE, 27502-85 Solutions and standards, and the 35613-21 meter.

      Answered on 7/7/2010 by Josh from Technical Support